Home ProductsPlastic Analyzer

Dichloromethane Soak Tester 1s Plastic Analyzer

Dichloromethane Soak Tester 1s Plastic Analyzer

    • Dichloromethane Soak Tester 1s Plastic Analyzer
  • Dichloromethane Soak Tester 1s Plastic Analyzer

    Product Details:

    Place of Origin: BEIJING, CHINA
    Brand Name: JINHAIHU
    Certification: ISO 9001
    Model Number: ELJWJZ-1

    Payment & Shipping Terms:

    Minimum Order Quantity: 1 set
    Packaging Details: Standard export wooden case
    Payment Terms: T/T, L/C, Western Union, MoneyGram, D/A, D/P
    Supply Ability: 10 sets per month
    Contact Now
    Detailed Product Description
    Product Name: Dichloromethane Soak Tester Model: ELJWJZ-1
    304# Stainless Steel Container Size: Φ200 Mm * 300 Mm High Methylene Chloride Thermostatic Temperature: 20±0.5 ℃
    Instrument Temperature Control Range: 10-50 ℃ The Precision Of The Instrument Shows: 0.1 ℃
    The Timer Range: 0-99 Minutes The Timer Precision: 1s
    High Light:

    dichloromethane soak tester

    ,

    dichloromethane soak analyzer

    ,

    1s plastic analyzer

    Overview

    The equipment is used for plastic pipe material, pipe fittings, sheet and immersion test for methylene chloride.Widely used in the manufacturer of pipe fitting, sheet, construction quality inspection test and research department.

     

    Standard

    GB/T13526: Rigid polyvinyl chloride (PVC -u) pipe methylene chloride immersion test method.

     

    Products are mainly composed of

    Constant temperature water bath tank, glass containers, sample fine strainer, temperature control and timing system.

     

    Technical indicators:

    304# stainless steel container size Φ200 mm * 300 mm high
    Methylene chloride thermostatic temperature 20±0.5 ℃
    Instrument temperature control range(with Refrigerating unit) 10-50 ℃
    The instrument temperature control precision ±0.2 ℃
    The precision of the instrument shows 0.1 ℃
    The timer range 0-99 minutes
    The timer precision 1s

    The sample standard size

    At least 160 mm sample intercept length

     

    Test method

    The depth should make sample submerged at least 30 mm
    2. Water seal layer thickness of about 20 mm
    3. Specimen constant temperature time 30 minutes, remove dry after 15 minutes.

    Dichloromethane Soak Tester 1s Plastic Analyzer 0

    Contact Details
    Beijing Jinshengxin Testing Machine Co., Ltd.

    Contact Person: sales

    Send your inquiry directly to us (0 / 3000)

    Other Products